中國臺灣華儀 7630 全功能接觸電流測試儀
7630 Touch Current Tester
7630 產(chǎn)品介紹
具有多項及創(chuàng)新功能,兼具大功率負載量測能力,適合所有接觸電流的測試應(yīng)用;超大型 320 x 240 Graphic LCD ,可同時顯示量測電壓/接觸電流及測試時間;完整的自動化接口,如網(wǎng)絡(luò)卡、USB & RS-232、IEEE-488 (GPIB) 及其他接口,可執(zhí)行自動化測試,亦可搭配華儀安規(guī)測試器,組成測試系統(tǒng),提升測試效率。
7630 產(chǎn)品特點
- 量測接觸電流時,亦可顯示人體仿真阻抗線路(MD)兩端的電壓 ,滿足接觸電流IEC 60990 標(biāo)準(zhǔn)要求
- 接觸電流可量測 AC/DC/AC+DC ,滿足醫(yī)療設(shè)備IEC 60601 標(biāo)準(zhǔn)要求
- 量測帶寬 DC - 1MHz,符合 IEC60990 接觸電流測試規(guī)范
- 可靈活選擇MD漏電流量測點(如: IEC60065 U1&U2),符合安規(guī)標(biāo)準(zhǔn)要求
- 內(nèi)建 7 組人體模擬阻抗(MD)與1組擴充空間,滿足多種安規(guī)標(biāo)準(zhǔn)要求
- 可搭配華儀安規(guī)測試器組成安規(guī)6HE一(ACW/DCW/IR/GB/RUN/LLT) 測試系統(tǒng),縮短測試時間
- MD 具有BNC量測端子,可外接示波器或電壓表讀取電壓值,方便對MD進行校準(zhǔn)
- 開機智能檢測功能,自動確認儀器輸出正常,確保測試的可靠度
- 校正及預(yù)警時間設(shè)定功能,提醒用戶定期校正,確保測試質(zhì)量
- 測試結(jié)果可儲存過程中大漏電流 (Imax),滿足客戶對質(zhì)量的要求
- 可選配大峰值 75mApeak / 有效值 35mArms 量測,滿足安規(guī)機構(gòu)測試驗證需求
中國臺灣華儀 7630 全功能接觸電流測試儀 主要技術(shù)參數(shù)
Specification : | ||
INPUT | ||
Voltage | 115/230 VAC ± 15% | |
Frequency | 50/60 Hz ± 5% | |
Fuse | 2A Slow-Blo 250V AC | |
Line condition | ||
Power Switch | Reverse polarity switch for normal condition (on / off / auto setting) | |
Neutral Switch | Neutral switch on/off selection for single fault condition | |
Ground Switch | Ground switch on/off selection for class I single fault condition | |
Probe Setting | ||
Surface to Surface (PH- PL) | ||
Surface to Line(PH - L) | ||
Ground to Line (G - L) | ||
Ground to Neutral (G - N) | ||
Auto Function (G - N & G - L) | ||
SETTING | ||
Touch Current High Limit (RMS) | Range: | 0.0uA~ 999.9uA/ 1000uA~5999uA/ 6.00mA~20.00mA |
Resolution: | 0.1uA/1uA/0.01mA | |
Touch Current Low Limit (RMS) | Range: | 0.0-999.9 uA |
Resolution: | 0.1uA | |
Touch Current High Limit (Peak) | Range: | 0.0uA - 999.9uA/ 1000uA - 9999uA/ 10.0mA - 30.0mA |
Resolution: | 0.1uA/1uA/0.1mA | |
Touch Current Low Limit (Peak) | Range: | 0.0-999.9 uA |
Resolution: | 0.1uA | |
DISPLAY | ||
Touch Current Display (RMS) | Range 1 | 0.0uA~ 999.9uA, frequency DC - 1MHz |
Resolution | 0.1uA | |
Accuracy | DC : ±2% of reading + 3counts | |
Range 2 | 1000uA~8500uA, frequency DC - 1MHz | |
Resolution | 1uA | |
Accuracy | DC : ±2% of reading + 3counts | |
Range 3 | 8.00mA~20.00mA, frequency DC - 1MHz | |
Resolution | 0.01mA | |
Accuracy | DC : ±2% of reading + 3counts | |
Touch Current Display (Peak) | Range 1 | 0.0uA~ 999.9uA, frequency DC - 1MHz |
Resolution | 0.1uA | |
Accuracy | DC : ±2% of reading +2uA | |
Range 2 | 1000uA ~ 8500uA, frequency DC - 1MHz | |
Resolution | 1uA | |
Accuracy | DC : ±2% of reading +2counts | |
Range 3 | 8.0mA ~ 30.0mA, frequency DC - 100KHz | |
Resolution | 0.01mA | |
Accuracy | DC : ±2% of reading +2counts | |
Measuring Device Module | ||
MD1 | IEC60990 Fig4 U2, IEC 60950-1, IEC60335-1, IEC60598-1, IEC60065, IEC61010 | |
IEC60990 Fig4 U1, IEC60065 | ||
MD2 | IEC60990 Fig5 U3, IEC60598-1 | |
IEC60990 Fig5 U1 | ||
MD3 | IEC 60601-1 | |
MD4 | UL544NP, UL484 , UL923, UL471, UL867, UL697 | |
MD5 | UL544P | |
MD6 | UL1563 | |
MD7 | IEC60950, IEC61010-1 FigA.2 (2K ohm) for RUN Test MD Circuit. | |
External MD | Basic measuring element 1k ohm | |
MD Voltage Limit | 70Vdc | |
Leakage Current Offset | Range: | 0 - 999.9uA |
Resolution: | 0.1uA | |
DUT POWER | ||
AC Voltage | 0.0-277.0V | |
AC Current | 40Arms max continuous | |
Over Current Protection | 50Arms, Response Time < 3 sec / 250Apeak Response Time <10u sec | |
AC Voltage High/Low Limit | Range: | 0.0-277.0V |
Resolution: | 0.1V/step | |
AC Voltage Display | Range: | 0.0 - 277.0 V |
Resolution: | 0.1 volt / step | |
Accuracy: | ±(1.5% of reading + 2 counts) , 30.0 - 277.0VAC | |
Delay time setting | Range: | 0.2 !V 999.9 seconds |
Resolution: | 0.1 second | |
Timer display | Range: | 0.0 !V 999.9 seconds |
Resolution: | 0.1 second | |
Accuracy: | ± (0.1% of reading + 0.05 seconds) | |
GENERAL | ||
PLC Remote Control | Input - Test, Reset, Interlock & Recall File 1 through10 | |
Output - Pass, Fail, Test-in-Process, Start-Out, Reset-Out | ||
Scanner Control | It is applied to control Matrix Scanner (Model 7006) | |
Memory | 50 memories , 30steps per each memory . Max. Result Display 900 data (30 memories x 30 steps) | |
Auto Reverse Function | AUTO Reverse ON / OFF parameter setting selection | |
Scope Output Interface | At rear panel BNC type to connet scope for some IEC standards test requirement and application. | |
Display | 320 X 240 graphic LCD | |
LCD Contrast Setting | Range: 1-9; 1 is lightest character, 9 is darkest character. | |
Alarm Volume Setting | Range: 0-9; 0=OFF, 1 is softest volume, 9 is loudest volume. | |
Security | Lockout capability to avoid unauthorized access to test set-up programs. | |
Calibration | Software and adjustments are made through front panel, | |
Results Display | All test result information will be displayed on the screen. | |
Interface | RS232 standard | |
Safety | Product with TUV/GS certificate | |
Dimension | 430 (W) x 133(H) x 300 (D) mm | |
Weight | 12 Kg | |
STANDARD ACCESSORIES | ||
Power Cord | X1 | |
Fuses | X2 (Including a spare contained in the fuse holder) | |
Interlock Disable Key(1505) | X1 | |
TC Receptacle Adaptor Box (1932) | X1 | |
Hipot Test Lead for PH / PL (1101) | X2 | |
RS232 Link Cable 1.5m(1130) | X1 | |
DUT Power Cable 40A(1148) | X1 | |
RUN Module (Opt.752) | ||
W | Range : 0-1KW / 1-10KW Resolution : 0.5W / 5W | |
A | Range : 0.000 !V 3.500 A / 3.00 - 40.00A , Resolution : 0.001A / 0.01A | |
PF | 0.000 -1.000 | |
Leakage Current | Range : 0.00 - 10.00mA , Resolution : 0.01mA | |
AC Current High/Low Limit Setting | Range: | 0-40A |
Resolution: | 0.01A/step | |
AC Current Display | Range: | 0.000 !V 3.500 A / 3.00 - 40.00A |
Resolution: | 0.001A / 0.01A | |
Accuracy : | ± (2 % of reading + 5 counts) / ± (2 % of reading + 2 counts) | |
AC Power High/Low Limit Setting | Range: | 0 - 10KW |
Resolution: | 1W | |
AC Power Display | Range: | 0.0 -1000.0W / 1000 - 10000W |
Resolution: | 0.5W / 5W | |
Accuracy : | ±(2 % of reading + 2 counts ) | |
Power Factor High/Low Limit Setting | Range: | 0.000 - 1.000 |
Resolution: | 0.001 | |
Power Factor Display | Range: | 0.000 - 1.000 |
Resolution: | 0.001 | |
Accuracy : | ± (2 % of reading + 2 counts ) | |
Leakage Current High/Low Limit Setting | Range: | 0.00 - 10.00mA |
Resolution: | 0.01mA | |
Leakage Current Display | Range: | 0.00 - 10.00mA |
Resolution: | 0.01mA | |
Accuracy : | ± (2 % of reading + 2 counts ) | |
Delay Timer Setting | Range: | 0, 0.2 - 999.9 |
Resolution: | 0.1 | |
Dwell Timer Setting | Range: | 0, 0.1-999.9 (0=continuous) |
Resolution: | 0.1 | |
Timer display | Range: | 0.0 - 999.9 seconds |
Resolution: | 0.1 second | |
Accuracy: | ± (0.1% of reading + 0.05 seconds) | |
Power Control | It is applied to control Transformer Box (1931/1931S) or AC Power Source(6700/6800/6400 Series) | |
HV & GB Link Module (Opt. 753) | ||
Max. ACW 4000VAC, DCW 4000VDC & Max. Ground Bond 40A | ||
GPIB Interface Card (Opt. 731) | ||
Printer Port Interface Card (Opt. 732) | ||
Multi-function Interface Card (Opt. 751) | ||
Including USB port A type, combination of RS-232, RS-485 port, and Bar Code Input with internal memory capability. | ||
High Measurement Range 35mArms / 70mApeak & 4MDs (Opt. 754 ) | ||
35mArms / 70mApeak for IEC60990 fig4(U1 and U2), IEC60990 fig5(U1 and U3), UL484, IEC60601 Test Requiremt , | ||
Transformer Box (Option) Add this option must optional Opt.752 RUN Module. | ||
1931 Series Transformer Box | Multi-Tape of 1.0, 1.06, 1.1 times of input voltage. | |
1931-S Series Transformer Box | Multi-Tape of 0.8, 0.85, 0.9, 1.0, 1.06, 1.1 times of input voltage. | |
AC Power Source (Option) Add this option must optional Opt.752 RUN Module. | ||
6800 Series Digital AC Power Source (6805, 6810 , 6820 , 6830 and 6840 ) | ||
6700 Series Programmable AC Power Source (6705, 6710 , 6720 , 6730 and 6740 ) | ||
6400 Series AC Power Source (6402, 6403, 6405, 6410 ) |