11S 氮氧化物檢測管 Nitrogen oxides
Nitrogen oxides 1S
NO+NO2
Performance |
Measuring Range | 5 to 10 ppm | 10 to 250 ppm | 250 to 625 ppm |
Number of PumpStrokes | 2 | 1 | 1/2 |
CorrectionFactor | 1/2 | 1 | 2.5 |
Sampling Time | 1 minute per pumpstroke | ||
Detecting Limit | 2 ppm (n=2) | ||
Color Change | White → Palegreen | ||
Reaction Principle | NO + Cr6++ H2SO4→NO2 | ||
Coefficient ofVariation | 10% (for 10 to 40ppm), 5% (for 40 to 250 ppm) | ||
ShelfLife | 2Years | ||
Correctionsfor temperature & humidity | Unnecessary | ||
Store the tubes at cool and darkplace. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Hydrogenchloride | 50 ppm | Faintdemarcation | Produce Bluishpurple |
Ozone | 80 ppm | Faintdemarcation | Pale brown |
Sulfur dioxide, | 1/1 | Plus error | Nodiscoloration |
Methanol | 400 ppm | Minus error | Nodiscoloration |
Othersubstance measurable with this detector tube |
Calibration gasgeneration | NO2:Permeation tube method NO:High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
NO2:3ppm NO:25ppm | NO2:5ppm | - |