醋酸乙烯酯檢測管143 Vinyl acetate
Vinyl acetate 43
CH3CO2CH:CH2
Performance |
Measuring Range | 5 to 10 ppm | 10 to 100 ppm | 100 to 250ppm |
Number of Pump Strokes | 4 | 2 | 1 |
Correction Factor | 1/2 | 1 | 2.5 |
SamplingTime | 45 seconds per pumpstroke | ||
DetectingLimit | 1 ppm (n=4) | ||
ColorChange | White → Brown | ||
ReactionPrinciple | CH3CO2CH:CH2+C6H3(CH3)3+ H2S2O7 → Condensation Polymer (Brown) | ||
Coefficient ofVariation | 15% (for10 to 20 ppm), 10% (for 20 to 100 ppm) | ||
ShelfLife | 3Years | ||
Corrections for temperature & humidity | Unnecessary | ||
Store the tubes at cool and darkplace. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Aldehydes | - | Plus error | Brown |
Ketones | - | Plus error | Brown |
Ethylene | 1000ppm | Plus error | Yellow |
Vinylchloride | 500ppm | Plus error | Yellow |
Othersubstance measurable with this detector tube |
Calibration gasgeneration | Diffusion tubemethod |
TLV-TWA | TLV-STEL | Explosive range |
10ppm | 15ppm | 2.6 to13.4% |