導(dǎo)熱系數(shù)測定儀是檢測儀器,用于測量各種勻質(zhì)板材狀絕熱保溫材料的導(dǎo)熱系數(shù)及復(fù)合板材的測定。使用前務(wù)必仔細閱讀說明書。并由實驗人員操作,以避免操作不當(dāng)引起的傷害。如需了解更多資料請與我公司客服人員聯(lián)系。
Thermal conductivity tester is a testing instrument for measuring the thermal conductivity of various homogeneous plate-like thermal insulation materials and the determination of composite plates. Be sure to read the instructions carefully before use. It is operated by experimental personnel to avoid injury caused by improper operation. For more information, please contact our customer service staff.
導(dǎo)熱系數(shù)測定儀按照GB/10294-2008《絕熱材料穩(wěn)態(tài)熱阻及有關(guān)特性的測定,熱流計法》設(shè)計制造。適用于檢測單位、大專院校、科研生產(chǎn)單位使用。采用大屏幕觸摸屏具有操作簡便,智能化程度高等優(yōu)點。
The thermal conductivity tester is designed and manufactured in accordance with GB / 10294-2008 "Determination of steady-state thermal resistance and related characteristics of thermal insulation materials, heat flow meter method". Applicable to testing units, colleges and universities, research and production units. The large-screen touch screen has the advantages of easy operation and high intelligence.
導(dǎo)熱系數(shù)測定儀主要特點:
★微機自動控制,自動完成數(shù)據(jù)采集和生成報表,智能化程度高,使檢測快捷、準確。
★主機箱采用復(fù)合沒計,使用方便,外觀新穎大方。
★冷單元部分采用自動汽缸夾緊,氣動系統(tǒng)壓力值可調(diào)。
★冷系統(tǒng)采用全封閉式壓縮機組,冷卻快、制冷均勻、運行噪音小。
★計量加熱單元采用紫銅板作為加熱面板。具有高導(dǎo)熱性,使溫度均勻熱慣性小,測量結(jié)果更準確。
Main characteristics of thermal conductivity tester:
★ Microcomputer automatic control, complete data collection and report generation automatically, high degree of intelligence, make detection fast and accurate.
★ The main box adopts composite design, convenient to use and novel in appearance.
★ The cold unit is clamped by an automatic cylinder, and the pressure value of the pneumatic system is adjustable.
★ The cooling system uses a fully enclosed compressor unit, which has fast cooling, uniform cooling and low operating noise.
★ The metering heating unit uses a copper plate as the heating panel. With high thermal conductivity, uniform temperature and low thermal inertia, more accurate measurement results.
導(dǎo)熱系數(shù)測定儀技術(shù)參數(shù):
★熱板溫度范圍:常溫—80℃
★冷板溫度范圍:10~50℃
★試件規(guī)格尺寸:300X 300mm
★試件厚度范圍:10~40mm
★測試準確度:≤±3%
★電源、功率:220V/2KW
★外形尺寸:800×600×1600mm
Technical parameters of thermal conductivity tester:
★ Temperature range of hot plate: normal temperature—80 ℃
★ Cold plate temperature range: 10 ~ 50 ℃
★ Specifications of test piece: 300X 300mm
★ Test piece thickness range: 10 ~ 40mm
★ Test accuracy: ≤ ± 3%
★ Power supply and power: 220V / 2KW
★ Dimensions: 800 × 600 × 1600mm
導(dǎo)熱系數(shù)測定儀工作原理:
型導(dǎo)熱系數(shù)測定儀,采用雙試件測定裝置,防護熱板組:包括加熱單元、冷卻單元。 加熱單元應(yīng)分為在中的計量單元和由隔縫分開的環(huán)繞計量單元的防護單元,并裝有絕熱裝置。 熱單元采用雙熱面加熱器,冷板與雙熱面對稱布置,根椐試件的厚度設(shè)定移動冷板的空間,將被測試件垂直放置在兩個相互平行具有恒定溫度的平板中,在穩(wěn)定狀態(tài)下,試件中測量部分具恒定熱流,通過測定穩(wěn)定狀態(tài)下流過計量單元的一維恒定熱流量Q、計量單元的面積A、試件冷、熱表面的溫度差T,可計算出試件的熱阻R,根據(jù)試件的厚度, 可以準確算出試件的導(dǎo)熱系數(shù)入值。
Thermal conductivity tester works:
Type thermal conductivity tester, using double test piece measuring device, protective hot plate group: including heating unit, cooling unit. The heating unit shall be divided into a metering unit in between and a protective unit surrounding the metering unit separated by a gap, and equipped with thermal insulation. The hot unit adopts double hot surface heaters. The cold plates are arranged symmetrically with the double hot surfaces. The thickness of the test piece sets the space for moving the cold plate. The test piece is placed vertically in two flat plates with constant temperature parallel to each other. In the steady state, the measurement part of the test piece has a constant heat flow. By measuring the one-dimensional constant heat flow Q flowing through the metering unit, the area A of the metering unit, and the temperature difference T between the cold and hot surfaces of the test piece, it can be calculated. According to the thickness of the test piece, the thermal resistance R of the test piece can be accurately calculated.
導(dǎo)熱系數(shù)測定儀注意事項:
接地保護線必須可靠接地,開機前檢查電源電壓是否正常。
制冷系統(tǒng)嚴禁連續(xù)啟動,啟動間隔應(yīng)大于3分鐘。
試件的測試溫度一定要低于試件的熔點溫度。
溫度過高,電源被切斷,查找原因后方可重新檢測。
試驗結(jié)束, 點 “自動檢測”鍵復(fù)位后再點“退出”鍵。
自動運行中不要點“退出” 和“手動調(diào)控”,否則實驗終止或出現(xiàn)失控。
制冷系統(tǒng)制冷劑為R十二。
系統(tǒng)通電時不許觸摸可能帶電部位。
本裝置在運輸搬運和移動位置時,禁止用桿子撬動底部,避免損壞 制冷系統(tǒng)。運輸搬運時利用頂部吊環(huán)移動或搬運。搬運時整機的傾斜角度不應(yīng)超過45°角,防止損壞制冷壓縮機。
Note for thermal conductivity tester:
The grounding protection wire must be reliably grounded. Check whether the power supply voltage is normal before starting up.
Refrigeration system is strictly forbidden to start continuously. The start interval should be more than 3 minutes.
The test temperature of the test piece must be lower than the melting point of the test piece.
The temperature is too high, the power is cut off, and the cause can be checked again.
After the test, click the "Auto Detect" button to reset and then click the "Exit" button.
Do not click "Exit" and "Manual Control" during automatic operation, otherwise the experiment will be terminated or out of control.
Refrigeration system refrigerant isR十二.
Do not touch any live parts while the system is powered on.
When the device is transported and moved, it is forbidden to pry the bottom with a pole to avoid damaging the refrigeration system. Use the top ring to move or carry during transportation. The tilt angle of the whole machine should not exceed 45 ° during transportation to prevent damage to the refrigeration compressor.