產(chǎn)品介紹:
Liquiline System CA80AM在所有關(guān)鍵控制點(diǎn)均可進(jìn)行高精度氨氮測(cè)量。作為L(zhǎng)iquiline平臺(tái)的一員,它可以即插即用,且易于升級(jí)至測(cè)量站 - **安裝需求。自動(dòng)標(biāo)定和清洗,以及低試劑消耗量可以節(jié)約運(yùn)作和維護(hù)成本。帶遠(yuǎn)程訪問(wèn)的**診斷功能確保了過(guò)程安全性,能夠向用戶(hù)提供過(guò)程文檔,方便過(guò)程管理。
方便調(diào)試,即插即用得益于系統(tǒng)采用Memosens技術(shù)以及友好的用戶(hù)操作界面
功能升級(jí)簡(jiǎn)便 - 即使升級(jí)至完整的測(cè)量站 - 僅需添加和連接Memosens傳感器
帶遠(yuǎn)程訪問(wèn)的**診斷功能,提升了過(guò)程安全性
自動(dòng)標(biāo)定,自動(dòng)清洗以及可選試劑冷卻單元,能夠有效降低系統(tǒng)維護(hù)及管理成本
符合標(biāo)準(zhǔn)比色法測(cè)量原理 - 靛藍(lán)法 - 符合ISO 7150-1、DIN 38406-5和GB 7481-87標(biāo)準(zhǔn)
Liquiline System CA80AM監(jiān)測(cè)和控制:
污水處理廠進(jìn)水口
污水處理廠出水口,用于文檔記錄
曝氣池
水處理
市政和工業(yè)污水處理廠的凈化能力
測(cè)量原理 | 比色法分析儀 | |
特點(diǎn) | Process analyzer for ammonium in aqueous solutions Comply with standard colorimetric measuring principle - indophenol blue method - following ISO 7150-1, DIN 38406-5 and GB 7481-87 | |
尺寸 | Cabinet 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Stand housing 1693 x 530 x 417 mm 66.65 x 20.87 x 16.42 in | |
設(shè)計(jì) | Open design, cabinet and stand housing High-Performance plastic ASA-PC, additional stand coated steel | |
過(guò)程溫度 | 4 ... 40 °C (41 ... 104 °F) | |
環(huán)境溫度 | 5 ... 40 °C (41 ... 104 °F) | |
過(guò)程壓力 | at atmospheric pressure, < 0.2 bar absolute | |
探頭處的流速 | min. 5ml/min (0,17 fl.oz/min) | |
探頭的一致性 | Suspended solids content Turbitity < 50 NTU, aqueous, homogenized | |
特殊要求 | Easy upgrade to measuring station with up to four digital Memosens sensors Automatic calibration and cleaning User-configurable measuring, cleaning and calibration intervals Optional cooling module for longer reagent shelf life 2 channel version optional User-definable measuring ranges Modular design for easily extensible functionality Digital communication for remote access | |
應(yīng)用 | Monitoring and optimization of the cleaning capacity of municipal and industrial wastewater treatment plants Monitoring of the wastewater treatment plant outlet for documentation purposes Monitoring and optimization of activated sludge basins | |
電源 | 100 ... 120 VAC / 200 ... 240 VAC ± 10% 24 VDC ± 10% 50 ± 1 oder 60 ± 1,2 Hz | |
輸出 | 2x 0/4 ... 20 mA Modbus RS485/TCP (optional) Ethernet (optional) Alarmrelay | |
輸入 | 1 or 2 measuring channel 1 ... 4 digitale sensor inputs for sensors with Memosens protocol (optional) | |
測(cè)量范圍 | 0,05 - 20 mg/l NH4-N 0,5 - 50 mg/l NH4-N 1 - 100 mg/l NH4-N | |
耗材 | Reagents and standard solutions CY80AM as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800. |