HV-P60A
HV-P60A
High-voltage Probe HV-P60A
HV-P60A
High-voltage waveforms of up to DC 60 kV, pulse 80 kV measured by oscilloscope
Wide bandwidth, DC to 50 MHz
Maintenance-free, due to use of silicon rubber - guarantees stable withstand voltage
Specifications | |
Maximum input voltage | DC 60 kV or AC peak* |
Bandwidth | DC to 50 MHz, within -3 dB |
Rise time | 7 ns or less (for probes only) |
Input RC | 1,000 MW ±5% //5 pF ±2 pF |
Attenuation | 2000:1 ±5% |
Cable length | Approx. 4 m |
Connector | BNC type |
Used for | Oscilloscope with input RC: 1 MW, 10 to 40 pF |
Weight | Approx. 3.2 kg (including accessories) |
Accessories | Matching box, clamp, probe tip, ground lead, adjusting screwdriver, accessory bag |
Options | High-voltage probe stand SK-301 (produced upon order) |
* The P-P voltage drops if the CW is 100 kHz or greater. |
This is a passive probe that measures high-voltage waveforms by connecting to an oscilloscope, and has the following characteristics: the bandwidth is DC to 50 MHz, the maximum input is DC 60 kV, and the attenuation is 2000:1.
Characteristic curve |
Maximum Input Voltage and Frequency (Sine Wave) [HV-P60A] |
High-voltage Probe HV-P30A
HV-P30A |
High-voltage waveforms of up to DC 30 kV, pulse 40 kV measured by oscilloscope | |
Wide bandwidth, DC to 50 MHz | |
Maintenance-free, due to use of silicon rubber - guarantees stable withstand voltage |
Specifications | |
Maximum input voltage | DC 30 kV or DC + AC peak* |
Bandwidth | DC to 50 MHz, within -3dB |
Rise time | 7 ns or less |
Input RC | 100 MW ±5% //7 pF or less |
Attenuation | 1000:1 ±5% |
Cable length | Approx. 3 m |
Connector | BNC type |
Used for | Oscilloscope with input RC: 1 MW, 10 to 40 pF |
Weight | Approx. 2.5 kg (including accessories) |
Accessories | Matching box, clamp, probe tip, ground lead, adjusting screwdriver, accessory bag |
Options | High-voltage probe stand SK-301 (produced upon order) |
* The P-P voltage drops if the CW is 100 kHz or greater. |
This is a passive probe that measures high-voltage waveforms by connecting to an oscilloscope, and has the following characteristics: the bandwidth is DC to 50 MHz, the maximum input is DC 30 kV, and the attenuation is 1000:1. |
Characteristic curve |
Maximum Input Voltage and Frequency (Sine Wave) [HV-P30A] |
Dimensions |