美國Cannon凱能BS/IP/MSL微型懸浮液位玻璃粘度計簡介:
The BS/IP/MSL Miniature Suspended Level Viscometer measures kinematic viscosity of transparent Newtonian liquids according to ASTM D445 and ISO 3104. Minimum sample volume is 4 mL. Minimum bath depth is 305 mm (12 in). Viscometers are provided with a certificate of calibration and instruction sheet. Holders are sold separately.
根據(jù)ASTM D 445和ISO 3104標(biāo)準(zhǔn),BS/IP/MSL微型懸浮液位粘度計測量了透明牛頓液體的運(yùn)動粘度。小進(jìn)樣量為4ml。小沐浴深度為305毫米(12英寸)。粘度計具有校準(zhǔn)證書和說明書。
美國Cannon凱能BS/IP/MSL微型懸浮液位玻璃粘度計技術(shù)參數(shù):
BS/IP/MSL Miniature Suspended Level Viscometers
Catalog Number | Size | Approximate Constant | Viscosity Range (mm²/s) | |
Low | High | |||
9724-G10 | 1 | 0.003 | 0.6 | 3.0 |
9724-G13 | 2 | 0.01 | 2.0 | 10 |
9724-G16 | 3 | 0.03 | 6.0 | 30 |
9724-G19 | 4 | 0.1 | 20 | 100 |
9724-G22 | 5 | 0.3 | 60 | 300 |
9724-G25 | 6 | 1.0 | 200 | 1000 |
9724-G28 | 7 | 3.0 | 600 | 3000 |
† Viscosity ranges in mm²/s represent flow times of 200 to 1000 seconds
美國Cannon凱能BS/IP/MSL微型懸浮液位粘度計方法介紹:
美國Cannon玻璃毛細(xì)粘度管測試方法:ASTM D445, ASTM D446, ASTM D2170方法,透明和不透明液體運(yùn)動粘度測定法 (包括計算動力粘度),玻璃毛細(xì)管運(yùn)動粘度計的規(guī)格及使用說明,瀝青運(yùn)動粘度的試驗方法
原理:本法用以測定透明與不透明產(chǎn)品的粘度,它是在已知的溫度下,利用液體本身的重力流動性(gravity-flow)來測定一定體積的液體通過經(jīng)校正后的毛細(xì)玻璃管粘度計所需的時間。根據(jù)所測出的動力粘度可計算出公制粘度或粘度(dynamic viscosity)。
方法概要:取一定體積的液體在嚴(yán)格的溫度與固定的液面高度的控制下,使其流經(jīng)毛細(xì)管粘度計而計算其流經(jīng)時間。根據(jù)流經(jīng)時間與粘度計的校正常數(shù)的乘積即可得動力粘度。
所需儀器:
1. 粘度計: 校正過后的玻璃毛細(xì)管,其所測得的粘度能符合于所要求的復(fù)驗準(zhǔn)確度再試準(zhǔn)確度。
2. 粘度計支架: 能將粘度計在恒溫槽中架持穩(wěn)固,且能使其垂直豎立。
3. 粘度計調(diào)溫槽: 內(nèi)盛裝透明液體,能使樣品于粘度計內(nèi)較液面低20mm以下,或高于槽底20mm以上,能保持槽內(nèi)溫度的變動于15至100℃間低于±0.02℃的變化,在此溫度范圍以外,亦不超過±0.05℃的溫度變化。
4. 測溫器: 檢驗動力粘度計的溫度計內(nèi)盛有適當(dāng)?shù)囊后w,使之能符合溫度范圍,溫度計使用前均需校正,校正后準(zhǔn)確度在±0.02℃,若其它測溫器有上述的精確度時亦可采用。
5. 定時器: 秒表或彈簧裝置的定時器均可使用,但其刻度至少在0.1秒,其測定200秒至900秒的誤差在±0.07%以內(nèi)。