1,1-二氯乙烯檢測(cè)管130L
Vinylidene chloride 30L
CH2:CCl2
Performance |
MeasuringRange | 0.4 to 1 ppm | 1 to 14 ppm | 14 to 40.6 ppm |
Number of Pump Strokes | 2 | 1 | 1/2 |
CorrectionFactor | 0.4 | 1 | 2.9 |
SamplingTime | 1 minute per pumpstroke | ||
DetectingLimit | 0.1 ppm (n=2) | ||
ColorChange | Yellow → Reddishbrown | ||
ReactionPrinciple | CH2:CCl2 + Cr6++ H2SO4 →HCl | ||
Coefficient ofVariation | 10% (for1 to 4 ppm), 5% (for 4 to 14 ppm) | ||
ShelfLife | 2Years | ||
Correctionsfor temperature & humidity | Unnecessary | ||
Store the tubes inthe refrigerator to keep at 10oC (50oF) orbelow. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Ethylene | 300ppm | Minus error | Nodiscoloration |
Vinyl chloride | 1/7 time | Plus error | Produce reddishbrown |
Tetrachloroethylene | 4 times | Plus error | Produce reddishbrown |
Trichloroethylene | 1time | Plus error | Produce reddishbrown |
Toluene | 500 ppm | Minus error | Nodiscoloration |
Benzene | 400 ppm | Minus error | Nodiscoloration |
Othersubstance measurable with this detector tube |
Calibration gasgeneration | Permeation tubemethod |
TLV-TWA | TLV-STEL | Explosive range |
5ppm | - | 5.6 to11.4% |