M-402 DH3000 Nucleic Acid Analyzer is a basic micro-volume UV spectrophotometer specifically designed (only three types of wavelength: 230nm, 260nm and 280nm) for the measurement of nucleic acids and proteins.
Features:
Flexible data output--The data can be printed with built-in printer, the data can output via SD-RAM Card and USB memory
Micro-volumes Easy & Fast--Only require 0.5~2μl samples to accurate determinate of nucleic acids, proteins
Long life & No consumables--Xenon flash lamp, 10 flashes, up to 10 years, no cells or cuvetes, no dilitions and expensive comsumables
No computer required--Touch screen and panel operation with built-in controls and software, without computer
Easy & Fast-- Direct, easy measurements within 5 sec No dilutions for samples and no warm up
Measure nucleic acid & protein--Measure nucleic acid concentration at 260nm, and purity using the 260/280 ratio. Measure purified protein concentration at 280nm
Model | M-402 DH3000 |
Wavelength Range | 230nm, 260nm,280nm |
Minimum Sample Size | 0.5~2.0ul |
Path Length | 0.2mm (For high concentration measurement) 1.0mm (For ordinary) |
Light Source | Xenon flash lamp |
Detector Type | Photodiode |
Absorbance Precision | 0.003 Abs |
Absorbance Accuracy | 1% (7.332Abs at 260 nm) |
Absorbance Range | 0.02 - 80 (10mm equivalent absorbance) |
Detects Nucleic Acid up to | 10~4000ng/ul (dsDNA) |
Measurement Time | < 5s |
Dimensions (W x D x H)mm | 210 x 280 x 166 |
Weight | 2.7kg |
Sample Pedestal Material | Aluminum alloy and Quartz fiber |
Operating Voltage | 24V DC |
Operating Power Consumption | 12 - 18 W |
Standby Power Consumption | 5W |